Quick answer: start from the device's load profile and decide whether the battery is the main supply or a backup. Size the runtime with a reserve for alarms and ageing. Give the device an accurate state of charge over SMBus or UART, and protect the pack with a BMS and a fuse. Keep spare batteries at or below 100 Wh where you can, because they ship more simply. Then lock the cell model, keep the batch traceability, and plan the battery certification alongside the device's own approval.
A battery in a medical device is judged differently from one in a tool or a scooter. A flat battery can interrupt a measurement or a therapy, and every change to it can mean retesting the device. This guide covers what device makers ask us most often. For our packs for medical and laboratory equipment, see medical device batteries; for small LiPo packs in detail, read LiPo packs for handheld, IoT and medical devices.
Main supply or backup?
The first question decides most of the others:
- Main supply: handheld instruments, portable monitors, wearable devices and transport equipment run from the battery for most of their use. The battery is sized for a full shift or mission, and charged between uses.
- Backup: mains-powered equipment, such as a bedside device, keeps running from the battery during a power cut or while it is moved between rooms. The battery spends most of its life fully charged and must carry the device for a defined period when the mains fails.
- Carts and workstations: mobile workstations and medical carts run from a larger battery for a shift and charge overnight or at a dock.
Write the requirement down as a number: for example, "8 hours of normal use plus 1 hour of reserve", or "30 minutes on battery after mains failure". It is the basis for every later decision.
Sizing runtime with a reserve
Take a portable monitor that averages 8 W and must run an 8-hour shift with a 1-hour reserve. These are example inputs; use the measured power of your device in its worst normal mode.
- Energy: 8 W × 9 h = 72 Wh.
- Margin for ageing, temperature and a pack that is never run completely flat: 72 ÷ 0.8 = 90 Wh.
- Our 10.8V 9Ah pack (3S2P, Molicel P45B) stores about 97 Wh and covers it, just under the 100 Wh transport threshold explained below.
The reserve matters more in medical equipment than elsewhere: the device must warn the user early enough to plug in or swap the battery, and that warning needs energy left in the pack. If the requirement grows to 12 hours, the same calculation gives 135 Wh. You can then choose one larger pack, such as the 10.8V 13.5Ah pack at about 146 Wh, or two hot-swappable packs of under 100 Wh each. For a mobile cart averaging 80 W over a 10-hour shift, 800 Wh ÷ 0.8 = 1 kWh points to the 25.6V 48Ah LiFePO4 pack, about 1.2 kWh. Our guide on how to size a battery pack covers the method in detail.
State of charge the device can trust
A user who sees "40%" must be able to rely on it. A simple protection board cannot tell the device how much energy is left; a smart BMS can. Our packs can be specified with SMBus, UART or another agreed interface, so the device reads state of charge, voltage, temperature and fault status and can show a battery gauge and trigger low-battery alarms. Two points help accuracy:
- Current counting, not voltage alone. Especially with LiFePO4, whose voltage stays flat over most of the discharge, state of charge is estimated from the current in and out, corrected at full charge.
- Agree the data early. Tell us which values your device firmware expects and how often, so the BMS and the device are designed together.
Read what a BMS does for the protection and communication functions.
Choosing the cell and format
- Handheld and wearable devices: LiPo pouch cells fill thin, flat spaces and come in many sizes. Our LiPo pouch cell range has more than 230 cells.
- Portable monitors, pumps and instruments: 18650 and 21700 lithium-ion cells in 2S to 4S layouts give a rigid, standard format and high energy per kilogram. See 18650 vs 21700.
- Carts and workstations: 25.6 V LiFePO4 packs are a common choice for their cycle life and thermal stability, where weight matters less; see LiFePO4 vs lithium-ion.
Whatever the format, every pack is built with named cells from recognised manufacturers, and you receive the datasheet and SDS/MSDS for the cells used.
Protection, charging and mechanical design
The BMS protects against over-charge, over-discharge, over-current and short circuit and balances the series groups. For medical devices, a fuse adds a second, independent layer of protection, and temperature sensors let the BMS stop charging when the pack is too cold or too hot. Lithium-ion and LiFePO4 cells should not be charged below about 0 °C unless the pack is designed for it, which matters for transport equipment and devices stored in cold vehicles.
Mechanically, the pack must fit the space in the device with clearance, with cells held firmly and connectors chosen for the number of insertions the device will see. Enclosures can be PVC heat-shrink inside the device housing, ABS hard plastic or steel. Our guide to battery pack enclosure design covers the details.
The 100 Wh line for shipping and service
Lithium-ion batteries up to 100 Wh can, under conditions, ship with simplified dangerous-goods requirements; above 100 Wh they are fully regulated Class 9 goods. A battery installed in or packed with the device ships as UN3481. For medical devices, spare batteries are shipped to hospitals and service centres for years, so a battery at or below 100 Wh can make logistics much simpler. Our guide to shipping lithium batteries explains the rules.

Traceability and change control
A medical device can be on the market for many years, and its battery needs to stay the same battery. Two practices help:
- Traceability: batches and cell date codes are traceable, and every shipment comes with a quality test report, so a field question can be traced back to the cells.
- Lock the cell model: once the battery is validated in your device, keep the cell model fixed. Changing the cell can mean retesting, so plan for long-term availability when you choose it. Our development workflow explains specification lock and documentation.
Certification: the battery and the device
Certification is reviewed per project, because requirements depend on the battery design, the device and the destination market. Typical examples for the battery are UN 38.3 for transport and IEC 62133 for safety. Regulatory approval of the medical device itself remains with the device manufacturer: the device is assessed under the medical-device rules of your market, typically including standards for medical electrical equipment such as IEC 60601-1, and the battery's documentation is one input to that assessment. Tell us the standards your device must meet at the start, because they affect the cell and the pack design. Read lithium battery certifications explained.
Packs to start from
| Pack | Layout | Energy | Typical use |
|---|---|---|---|
| 7.4V 6.8Ah LiPo | 2S2P pouch | 50.3 Wh | handhelds, scanners, small instruments |
| 14.8V 4.4Ah LiPo | 4S2P pouch | 65.1 Wh | portable equipment with higher power |
| 10.8V 7Ah Li-ion | 3S2P, Samsung 35E | 76 Wh | portable monitors and pumps |
| 10.8V 9Ah Li-ion | 3S2P, Molicel P45B | 97 Wh | portable devices, just under 100 Wh |
| 10.8V 13.5Ah Li-ion | 3S3P, Molicel P45B | 146 Wh | longer runtime, above 100 Wh |
| 25.6V 24Ah LiFePO4 | 8S4P, IFR32700 | 614 Wh | medical carts and workstations |
| 25.6V 48Ah LiFePO4 | 8S8P, IFR32700 | 1.2 kWh | carts with longer shifts |
Energy from the product pages. Every pack is built to order: voltage, capacity, BMS interface, connector and enclosure are fitted to your device, and prototypes have no minimum order. See also the 25.6V LiFePO4 batteries for cart applications.
Getting a medical device battery
Send us the device's voltage range, its average and peak current, the runtime and reserve you need, the space available, the state-of-charge interface and the standards the device must meet. Our engineers will propose the cell, the layout, the BMS and the documentation. Request a quote or see custom battery packs for how we work.
Frequently asked questions
What battery chemistry is used in medical devices?
Portable and handheld devices mostly use lithium-ion 18650 or 21700 cells or LiPo pouch cells for their energy density. Medical carts and workstations often use LiFePO4 for its cycle life and thermal stability. The right choice depends on the load, runtime, size and weight of the device.
Are your battery packs certified for medical devices?
Certification is arranged per project, where the design, application and destination market require it; typical examples for the battery are UN 38.3 and IEC 62133. Regulatory approval of the medical device itself remains with the device manufacturer.
Can the battery report its state of charge to the device?
Yes. The BMS can be specified with SMBus, UART or another agreed interface, so the device can read state of charge and battery status and warn the user before the battery runs low.
Why does 100 Wh matter for a medical device battery?
Lithium-ion batteries up to 100 Wh can, under conditions, be shipped with simplified dangerous-goods requirements; above 100 Wh they are fully regulated Class 9. Staying at or below 100 Wh per battery can simplify spare-battery shipping and service.
What documentation comes with a medical battery pack?
A complete battery datasheet once the specification is locked, the datasheet and SDS/MSDS for the cells used, and a quality test report with every shipment. Batches and cell date codes are traceable.